Four, five and
six-panel drug configurations offered
Photocopy template
included for recording results
Easy to read results
– no machine required
Drug Name:
Abbr:
Cutoff:
Amphetamine
AMP
50ng/mL
Methamphetamine
mAMP
50ng/mL
Marijuana
THC
12ng/mL
Phencyclidine
PCP
10ng/mL
Cocaine
COC
20ng/mL
Opiates
OPI
40ng/mL
Instructions
Section
I: Sample Collection
Instruct the donor not to place
anything in the mouth including
food, drink, gum or tobacco products
for at least 10 minutes prior to
collection.
Remove
Sample Collector Protector.
Insert Sponge end of collector
into mouth.
Actively swab the inside of the
mouth and tongue with Sponge. As
soon as the Sponge softens
slightly, gently press the
Sponge between the tongue and
teeth to ensure complete
saturation. The Sponge
is saturated when no hard spots
can be felt. Collect for at
least three (3) minutes before
removing Sponge.
Hand the saturated collector
over to the tester.
Section
II: Testing
A.
Align the Red Arrow on the
device with either one of
the White Marks
on the Cap. Insert the
collector vertically
into the Cap and
press down firmly until the
Cap reaches the thread.
B.
Twist the Handle clockwise
180° to tighten the Cap
until the Red Arrow
lines up with the other
White Mark.
C.
Place the test device
horizontally on a clean and
level surface with result
window facing up.
Read
results at five (5) minutes.
Results stable up to sixty (60)
minutes.
* All positive
results are presumptive and should
be confirmed by an alternative
method. The preferred confirmation
method is GC/MS or GC/MS/MS.
If
results are positive, secure Cap
with Security Seal and send to
laboratory for confirmation.